PolyRange Function (application for single-range balances)
The PolyRange function divides a single weighing range into as many as
4 ranges, each with a different readability. In the various ranges, the readability
will adjust so that the last numeral of a weight readout is displayed with
a resolution of 1, 2, 5 or 10 digits (10 digits = only the next to the last numeral
of the readout will change; the last numeral is blanked).
The PolyRange function makes filling easier because the readability becomes
slightly coarser as the load increases and you will not immediately lose an entire
place of readability.
Press the tare key at any range level to restore the full resolution of the first range,
even when the balance is loaded.
Code
Display mode 1st range 2nd range 3rd range
Highest possible accuracy 1 8 1* 3 2 1* 3 4 1*
Last numeral blanked when
load changes 1 8 2* 3 2 2 3 4 2*
Rounding factor 2 1 8 3 3 2 3 3 4 3
Rounding factor 5 1 8 4 3 2 4 3 4 4
Rounding factor 10 1 8 5 3 2 5 3 4 5
1.0% accuracy 1 8 6 3 2 6 3 4 6
0.5% accuracy 1 8 7 3 2 7 3 4 7
0.2% accuracy 1 8 8 3 2 8 3 4 8
0.1% accuracy 1 8 9 3 2 9 3 4 9
0.05% accuracy 1 8 10 3 2 10 3 4 10
0.02% accuracy 1 8 11 3 2 11 3 4 11
0.01% accuracy 1 8 12 3 2 12* 3 4 12
PolyRange function 1 8 13 3 2 13 3 4 13
* = factory setting, depends on the balance model in some cases