Counting with Additional Requirements
Changing the Reference Sample Quantity
You can change the reference sample quantity as required:
– Enter the reference sample quantity (piece count) using the numeric keys and store
it with F identified by “STO REF”
or
– Change it in the parameter mode
To select the parameter mode: – Hold down i for more than 2 seconds
To change the setting: – Use f identified by “^”
If you enter a wrong number, press c to clear it
To store the reference quantity
– temporarily until you turn off
the power: – briefly press STO - F
– permanently after you turn off
the power (in the non-volatile
memory): – press STO - F for more than 2 seconds
This setting is not cancelled by the reset code 9 – – 1°!
Reference Sample Quantity
Changing the reference sample quantity Code
No change 2 3 1
In cycles: 5,10, 20, 50, 100, 5,10, ... 2 3 2*
In increments of one 2 3 3
Storage Parameter for the Reference Sample Weight
(Code only for models MC 210 S (-0CE), MC 210 P (-0CE), MC 410 S (-0CE))
The weight is stored Code
with full accuracy according to the internal resolution 3 5 1
according to the display accuracy 3 5 2*
* = factory setting
3–58