Directed Electronics Automate 210D Instructions d'exploitation Page 138

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 188
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 137
Readout Meaning Reason/Solution
Err 80 Arithmetic error occurred Do not try to store “0” as
during the density the mass value.
determination program
Err 234 Initial activation of the Turn the balance off and back
“differential weighing on again.
and backweighing program”
The rechargeable battery for Leave the balance connected
the “differential weighing to line current for approx. 12 hours
and backweighing” memory to charge the battery.
(RAM) is decharged or If this error code continues
defective. to be displayed, contact your local
Sartorius Service Center.
3–60
Vue de la page 137
1 2 ... 133 134 135 136 137 138 139 140 141 142 143 ... 187 188

Commentaires sur ces manuels

Pas de commentaire