Post-Run Procedures
106
Part # 15035786 Rev. D
Switch Sequencing Modes
Use the Mode Select command from the Welcome screen to switch between high output
modes and Rapid Run mode.
Only runs of the same mode can be performed simultaneously. Therefore, mode changes
are applied to both flow cell A and flow cell B. If either flow cell is in progress, a mode
change is not possible.
A maintenance wash and gasket change are required when switching between run modes.
For more information, see Perform a Maintenance Wash on page 100.
Switch from High Output to Rapid Run Mode
Switching from a high output mode (HiSeq v4 or TruSeq v3) to Rapid Run mode requires a
rapid-mode maintenance wash.
Specification Rapid Maintenance Wash
Flow Cell Type
Rapid flow cell (2 lanes)
Flow Cell Gasket
10-port gasket and 8-port gasket
Reagents
Tween 20 and ProClin 300
Expected Volumes (ml)
60.2 ml
Time (minutes)
60 minutes
Switch from Rapid Run to High Output Mode
Switching from Rapid Run mode to a high output mode (HiSeq v4 or TruSeq v3) requires a
rapid-mode maintenance wash followed by high-output maintenance wash.