Directed Electronics 7111A Spécifications Page 38

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 46
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 37
PN 718113AF
26
Laser Induced Fluorescence Detection
Operation
Once the CCF value is calculated and accepted by the user, all subsequent measurements will be
multiplied by this value.
As long as run conditions, capillary I.D. and chemistries remain unchanged, the CCF calculated
should remain valid. It is recommended that the standard solution be measured (as a check)
periodically or when a new capillary is installed.
NOTE
The Calibration Correction Factor determined in this process will affect all subsequent analysis.
CAUTION
The LIF System must be installed and lasers must be warmed up according to the
laser manufacturer’s instructions before running the Calibration Wizard.
Calibration Wizard
To access the Calibration Wizard
1
Right-click on the LIF instrument from the 32 Karat home screen.
2
Choose Configure > Instrument and click the Configure button.
3
Double-click on the LIF Detector icon in the Configured modules window.
Vue de la page 37
1 2 ... 33 34 35 36 37 38 39 40 41 42 43 44 45 46

Commentaires sur ces manuels

Pas de commentaire