Laser Induced Fluorescence Detection
Operation
Once the CCF value is calculated and accepted by the user, all subsequent measurements will be
multiplied by this value.
As long as run conditions, capillary I.D. and chemistries remain unchanged, the CCF calculated
should remain valid. It is recommended that the standard solution be measured (as a check)
periodically or when a new capillary is installed.
NOTE
The Calibration Correction Factor determined in this process will affect all subsequent analysis.
CAUTION
The LIF System must be installed and lasers must be warmed up according to the
laser manufacturer’s instructions before running the Calibration Wizard.

Calibration Wizard

To access the Calibration Wizard
1
Right-click on the LIF instrument from the 32 Karat home screen.
2
Choose
Configure > Instrument
and click the
Configure
button.
3
Double-click on the LIF Detector icon in the Configured modules window.