PN 718113AF
20
Laser Induced Fluorescence Detection
Operation
3
Choose the type of instrument you want to control, P/ACE MDQ CE or ProteomeLab PA 800.
A warning notice appears stating that changing the instrument type will erase ALL device
settings. Since this is a new instrument configuration and there is no data to lose, choose
OK to
continue. Then click
Configure.
Figure 13 Confirmation
There are two options for configuring an instrument, manual or auto configure. Manual
configuration enables the user to customize the instrument configuration. Auto Configuration
will detect the type of detector installed and then scan the trays to determine what is installed.
The instrument needs to be powered on with the desired trays installed for Auto Configure to
work.
4
Double-click on the LIF Detector on the right side of P/ACE MDQ CE Configuration Screen.
See Figure 14.
Figure 14 Choosing the LIF Detector